The electrosurgical device is controlled from the built-in microcontroller that sets all the operation modes and provides storage of the previous operation modes and power levels.
The present model operates in mono- and bipolar modes, has sufficiently high power (no less than 200 W) and a developed alarm system, which prevents injury of a patient or a surgeon during endosurgical interventions.
Technical data
|
Power supply
|
220 V, 50 Hz
|
|
Major generation frequencies
|
440 ±11 kHz
|
|
Nominal device output power with the
nominal load of 500 Om -cutting mode
-coagulation mode
|
220± 44 W
100 ±20 W
|
|
Nominal device output power with the nominal load of 500 Om in a bipolar coagulation mode
|
90 ±18 W
|
|
Power consumption
|
no more than 650 W
|
|
Dimensions and weight of the device
|
320х335х140 mm; 7,0 kg
|
|
|